亚洲狼人第一精品影视|日韩欧美国产吃瓜欧美激情第21页|五月天激情综合日本ji91av|777天堂麻豆爱综合视频

歡迎來到匠仁醫療設備有限公司網站!

更多 > 新聞資訊
公司新聞
行業知識
檢測答疑
圖標 首頁 >> 新聞資訊 > 行業知識

行業知識

硬性內窺鏡如何正確保養?

來源:http://www.dcgr.cn/ 發布時間:2021-08-13 瀏覽量:0

硬性內窺鏡在使用后的保養程度,會直接影響硬鏡的壽命。只有認真地做好日常保養,才能夠延長硬鏡的使用壽命、減少硬鏡的磨損和老化、降低維修成本、減少不必要的停機時間。
The maintenance degree of rigid endoscope after use will directly affect the service life of rigid endoscope. Only through careful daily maintenance can we prolong the service life of the hard mirror, reduce the wear and aging of the hard mirror, reduce the maintenance cost and reduce unnecessary downtime.
(1)手術結束后,應進行床旁檢查。確認硬鏡使用過程在有無出現故障,如外力導致插入部彎曲變形;插入部是否出現被利器刺傷;目鏡部有無破損情況;物鏡部是否出現蓋玻璃破損或光導纖維先端部磨損等情況。如果出現問題,應立即停止后續清洗步驟,防止硬鏡故障擴大。
(1) After the operation, bedside examination should be carried out first. Confirm whether there is any fault during the use of the hard mirror, such as bending and deformation of the insertion part caused by external force; Whether the insertion part is stabbed by sharp tools; Whether the eyepiece is damaged; Whether the cover glass of the objective lens is damaged or the tip of the optical waveguide is worn. In case of any problem, immediately stop the subsequent cleaning and disinfection steps to prevent the expansion of hard mirror fault.
(2)床旁檢查結束后,才能對硬鏡鏡體進行清洗。待導光接頭自然冷卻后,應盡快將鏡體沖洗干凈;物鏡鏡頭前端和有污漬時應盡快用紗布擦拭干凈,防止污漬干涸后影響光路或圖像通路。器械部如果有可拆卸的組件,應拆卸下來仔細清洗干燥,防止生銹。清洗完后進行干燥,并將硬鏡妥善存放在專用的內鏡器械盒內,準備進行。
(2) After the bedside inspection, the hard mirror body can be cleaned. After the light guide joint cools naturally, wash the mirror body as soon as possible; When there are stains on the front end and of the objective lens, wipe it with gauze as soon as possible to prevent the light path or image path from being affected after the stains dry up. If there are detachable components in the instrument department, they should be disassembled, cleaned and dried carefully to prevent rust. After cleaning, it shall be fully dried, and the hard mirror shall be properly stored in a special endoscopic instrument box for sterilization.
以上就是匠仁醫療的小編為大家介紹關于內窺鏡維修的介紹,希望可以幫助到您,更多精彩內容可以直接關注:http://www.dcgr.cn
The above is the introduction about endoscope maintenance introduced by the editor of Jiangren medical. I hope it can help you. More wonderful contents can be directly focused on: http://www.dcgr.cn
s
新聞資訊
聯系我們
  • 公司公司:匠仁醫療設備有限公司
  • 匠仁醫療山東濟南維修部
  • 聯系人:樊經理
  • 電話:13153199508
  • 地址:山東省濟南市槐蔭區美里東路3000號德邁國際信息產業園6號樓101-2室
  • 匠仁醫療湖南長沙維修部
  • 聯系人:李經理
  • 電話:13873135765
  • 地址:湖南省長沙市雨花區勞動東路820號恒大綠洲14棟2409室
新聞資訊

樊經理:13153199508     李經理:13873135765

公司地址:山東省濟南市槐蔭區美里東路3000號德邁國際信息產業園6號樓101-2室  湖南省長沙市雨花區勞動東路820號恒大綠洲14棟2409室

     

魯公網安備 37010402441281號