亚洲狼人第一精品影视|日韩欧美国产吃瓜欧美激情第21页|五月天激情综合日本ji91av|777天堂麻豆爱综合视频

歡迎來(lái)到匠仁醫(yī)療設(shè)備有限公司網(wǎng)站!

更多 > 新聞資訊
公司新聞
行業(yè)知識(shí)
檢測(cè)答疑
圖標(biāo) 首頁(yè) >> 新聞資訊 > 檢測(cè)答疑

檢測(cè)答疑

內(nèi)窺鏡的故障出在哪你知道嗎?

來(lái)源:http://www.dcgr.cn/ 發(fā)布時(shí)間:2021-11-30 瀏覽量:0

內(nèi)窺鏡分為硬鏡和軟鏡,硬鏡,一般使用在外科領(lǐng)域,插入管大多由不銹鋼材料組成,也有前部可彎曲的硬鏡,包括腹腔鏡、鼻竇鏡,前列腺電切鏡、輸尿管鏡及宮腔鏡等 。
Endoscope is divided into hard mirror and soft mirror, hard mirror, commonly used in surgical field, insert tube mostly by stainless steel material composition, also have front can bend hard mirror, include celioscope, nose sinuses mirror, prostate electrocision mirror, ureteroscope and hysteroscope to wait.
軟鏡,內(nèi)科領(lǐng)域比較多見(jiàn),如腸胃鏡、十二指腸鏡、喉鏡、氣管鏡、插管鏡及輸尿管腎盂鏡等。
Soft endoscopy, medical fields more common, such as gastroscope, duodenoscope, laryngoscope, tracheoscope, intubation and Ureteropelvic Mirror and so on.
但是,在使用的過(guò)程中,內(nèi)窺鏡難免會(huì)出現(xiàn)這樣那樣的故障,一般它的故障出在哪兒,今天就讓匠仁醫(yī)療內(nèi)窺鏡設(shè)備維修的小團(tuán)隊(duì)來(lái)給大家說(shuō)一下。
However, in the use of the process, the endoscope will inevitably have this kind of failure, it is generally where the failure, today let artisan medical endoscope equipment repair team to tell you.
一般故障會(huì)出現(xiàn)以下幾個(gè)方面:
Common failures include the following:
顯示器沒(méi)有圖像顯示
為了確保圖象主機(jī)和顯示器已接通,圖象主機(jī)和顯示器的電源開(kāi)關(guān)指示燈是否亮,如不亮,則需進(jìn)行保險(xiǎn)及供電檢查,檢查圖象主機(jī)與顯示器的連線、 IN和 OUT接口是否連接。在具有多種視頻信號(hào)源選擇器的顯示器中,選擇合適的視頻信號(hào)。查看顯示器的相應(yīng)設(shè)置參數(shù),將亮度、顏色、對(duì)比度默認(rèn)設(shè)為默認(rèn),確保攝像機(jī)的連接電纜固定在圖像主機(jī)上。在圖象主機(jī)和顯示器之間更換視頻電纜,排除可能的電纜損壞。
Display no image display in order to ensure that the image host and the display are connected, whether the power switch indicator light of the image host and the display is on or not, if it is not, a power supply check should be carried out, check the connection between the image host and the Monitor, and the IN and OUT interfaces. In a display with a variety of video signal source selectors, select the appropriate video signal. Check the display settings, set the brightness, color, contrast to the default, to ensure that the camera cable fixed in the image host. Replace the video cable between the image host and the Monitor to eliminate possible cable damage.
監(jiān)視
錯(cuò)誤的圖像顏色
錯(cuò)誤的圖象顏色要求 WB重新校正。先確定 RGB或 YPbPr接口沒(méi)有接線錯(cuò)誤,顯示參數(shù)確定,主機(jī)參數(shù)確認(rèn),選擇其它視頻通道進(jìn)行測(cè)試,進(jìn)行 COLOR BAR測(cè)試。檢查設(shè)定菜單的字體和顏色是否正常,如果正常,但內(nèi)視鏡圖像不正常,基本上是內(nèi)視鏡或攝像頭故障。
Monitor the wrong image color the wrong image color requires WB recalibration. First determine RGB or YPbPr interface no wiring error, display parameters determined, host parameters confirmed, select other video channels for testing, COLOR BAR testing. Check that the font and color of the settings menu are normal, if so, but the endoscopic image is not normal, basically an endoscopic or camera failure.
圖象干擾
發(fā)生干擾時(shí)應(yīng)檢查使用的高頻、強(qiáng)電的接地是否有問(wèn)題,外部設(shè)備的信號(hào)線是否有問(wèn)題。
If image interference occurs, check the grounding of the high frequency and high current used and the signal line of the peripheral equipment.
好了今天匠仁醫(yī)療內(nèi)窺鏡設(shè)備維修給大家先介紹到這里,相信你看過(guò)后,以后會(huì)注意內(nèi)窺鏡的使用和維護(hù)了吧,接下來(lái)關(guān)于內(nèi)窺鏡的其它問(wèn)題,關(guān)注官網(wǎng)http://www.dcgr.cn/,我們繼續(xù)更新。
Okay, today, artisan medical endoscopic equipment maintenance to everyone first introduced here, I believe that after you have seen, you will pay attention to the use and maintenance of the endoscope, right? Next on the other issues of the endoscope, attention to the official website, we continue to update.
 
 
s
新聞資訊
聯(lián)系我們
  • 公司公司:匠仁醫(yī)療設(shè)備有限公司
  • 匠仁醫(yī)療山東濟(jì)南維修部
  • 聯(lián)系人:樊經(jīng)理
  • 電話:13153199508
  • 地址:山東省濟(jì)南市槐蔭區(qū)美里東路3000號(hào)德邁國(guó)際信息產(chǎn)業(yè)園6號(hào)樓101-2室
  • 匠仁醫(yī)療湖南長(zhǎng)沙維修部
  • 聯(lián)系人:李經(jīng)理
  • 電話:13873135765
  • 地址:湖南省長(zhǎng)沙市雨花區(qū)勞動(dòng)?xùn)|路820號(hào)恒大綠洲14棟2409室
新聞資訊

樊經(jīng)理:13153199508     李經(jīng)理:13873135765

公司地址:山東省濟(jì)南市槐蔭區(qū)美里東路3000號(hào)德邁國(guó)際信息產(chǎn)業(yè)園6號(hào)樓101-2室  湖南省長(zhǎng)沙市雨花區(qū)勞動(dòng)?xùn)|路820號(hào)恒大綠洲14棟2409室

     

魯公網(wǎng)安備 37010402441281號(hào)